The main task of technical committees is to prepare International Standards. Draft International Standards adopted by the technical committees are circulated to. à aiguille interne. STANDARD. ISO. Second edition. Reference number. ISO (E). Provläsningsexemplar / Preview. STANDARD. ISO. Second edition. Reference number. ISO (E). This is a free 4 page sample. Access the full version online.

Author: Vikasa Duzilkree
Country: Uruguay
Language: English (Spanish)
Genre: Politics
Published (Last): 15 July 2010
Pages: 211
PDF File Size: 3.19 Mb
ePub File Size: 13.79 Mb
ISBN: 332-5-69210-874-2
Downloads: 87600
Price: Free* [*Free Regsitration Required]
Uploader: Shagal

If a steel tube is used, it shall connply with ISO Fixed or removable fitting per- mitting venting of air while restricting or preferably preventing the escape of blood. E-5 Test report The test report shall contain at least the following information: Bureau of Indian Standards BIS is a statutory institution established under the Bureau of Indian Standards Act, to promote harmonious development of the activities of standardization, marking and quality certification of goods and attending to connected matters in the country.

Until that time, a manufacturer may label his product “radio- opaque” provided he can support this claim by demonstrat- ing that he has an appropriate method for showing radio- opacity.

EN ISO – European Standards

Catheter designed for the introduction or withdrawal of liquids or devices into or from the peripheral vascular system. Attention is particularly drawn to the following: The details of conditions under which the licence for the use of the Standard Mark may lso granted to manufacturers or producers may be obtained from the Bureau of Indian Standards.

Measure the time taken for fluid to form the first falling drop at the back of the vent fitting. Users of Indian Standards should ascertain that they are in possession of the latest amendments or edition by referring to the latest issue of ‘BIS Catalogue’ and ‘Standards: The catheter tube may have a single lumen or multiple lumens.

I Principle Water is allowed to flow through the catheter and the amount of flow is measured either volumetrically or gravimetrically.

The fluid is allowed to flow into the needle, and the time taken for fluid to leak through the vent fitting is measured. Standards are also reviewed periodically; is standard alongwith amendments is reaffirmed when such review indicates that no changes are needed; if the review indicates that changes are needed, it is taken up for revision. Page 3, subclause 4.

  GROKKING GIMP PDF

For the purpose of deciding whether a particular requirement of this standard is complied with, the final value, observed or calculated expressing the result of a test or analysis, shall be rounded off in accordance with IS 2: Pages 1 and 2 4. Peripheral intravascular catheter designed for the introduction of withdrawal of liquids or devices into or from the peripheral vascular system.

In this adopted standard, reference appears to certain International Standards for which Indian Standards also exist. The fluid pathway in the needle shall be free of unintended obstructions that would prevent flashback. Details of which is given in National Annex A.

Search the history of over billion web pages on the Internet. This does not preclude 10555–5 free use, in the course of implementing the standard, of necessary details, such as symbols and sizes, type or grade designations.

NOTE To improve the visibility of the solution, a colorant such as red or blue food dye may be incorporated.

When tested in accordance with annex E, fluid shall not 105555-5 out of the vent fitting within 15 s. Assembly comprising the cath- eter tube, catheter hub and any integral fittings. Add the following reference: This Indian Standard has been developed from Doc No.: Designation and catalogue of CCR colours.

Round the calculated average water fiowrate to the 105555-5 whole number of millilitres. Blood flow into the needle hub. No part of the these publications may be reproduced in any form without the prior permission in writing of BIS. The corresponding Indian Standards which are to be substituted in their places are listed below along with their degree of equivalence for the editions indicated: Certain conventions are, however, not isk to those used in Indian Standards.

Figure 1 — Typical over-needle peripheral intravascular catheter NOTE 2 At the time of publication of this part of ISOthere is no acceptable, validated test method to determine radio-detectability. Part 1 General requirements do The technical committee has reviewed the provisions of the following International Standard referred in this adopted standard and has decided that it is acceptable for use in conjunction with this standard: I Principle A force is applied successively tensile and connpress- ive to the needle tube and needle hub and the tube- hub union is then examined for loosening.

  ERNEST HEMINGWAY POR QUIEN DOBLAN LAS CAMPANAS PDF

An example of a suitable apparatus is shown in fig- ure B. The number of significant places retained in the rounded off value should be the 10555–5 as that of the specified value in this standard. Suggested colour references for opaque naaterials are given in annex C. International Standard ISO NOTE 3 The point should be designed to be noncoring. Review of Indian Standards Amendments are issued to standards as the need arises on the basis of comments.

ISO 10555-5:1996

Copyright BIS has the copyright of all its publications. An approved test method for producing a value of radio-detectability will be established. Assembly comprising at least a needle tube attached to, and communicating with, a needle hub.

10555- the time of publication, the editions indicated were valid. Fitting attached to the needle tube, providing communication with its bore.

Available from Munsell Color, N. When the needle is fully inserted into the catheter unit, the catheter tube shall neither extend beyond the heel of the needle bevel nor be nnore than 1 nnnn fronn it see dinnension a in figure 1.

Limited collection of colours. Rigid tube with one end sharpened to facilitate entry into body tissue.

Other parts of this standard are as follows: All standards are subject to revision, and parties to agreements based on this part of ISO are encouraged to investigate the possibility of applying the most recent editions of the standards indicated below. The assembly 105555-5 at least a needle tube attached to and communicating with a needle hub.

Annex D shows examples of typical needle point geom- etries.